Section 2 Getting Started
2.7
2.3.6 Serial Handshake Selection
Switch SW1-5 is used to select between hardware [
RTS/CTS
] or
software [
XON/XOFF
] serial handshake control.
With
XON/XOFF
, the Micro488/EX issues an
XOFF
character [ASCII
value of
&H13
] when its buffer memory is near full. When issued, there are
greater than 1000 character locations remaining to protect against buffer
overrun. When it is able to accept more information it issues an
XON
character [ASCII value of
&H11
]. With this handshake, the Micro488/EX
issues an
XON
character at reset or power-on. The Micro488/EX also
accepts
XON/XOFF
on transmit from the serial host it is communicating
with.
RTS/CTS
serial control becomes inactive when
XON/XOFF
is
enabled. The
RTS
output is, however, set to an active high state. The
CTS
input is not used for this handshake and may be left floating (unconnected).
With
RTS/CTS
, the Micro488/EX un-asserts
RTS
(low) when its
buffer memory is near full. When un-asserted, there are greater than 1000
character locations remaining to protect against buffer overrun. When it is
able to accept more information it asserts (high)
RTS
. The Micro488/EX
will not transmit data to the serial host if it detects the
CTS
input un-
asserted (low) when configured for this hardware handshake.
The factory default serial control is hardware,
RTS/CTS
.
SW1 View for Serial Handshake Selection
OPEN