CB-7012, CB-7014, CB-7017 Users Manual
2.18 ~AAEV
Description: Enable/Disable Calibration
Syntax: ~AAEV[CHK](cr)
~ Delimiter character
AA Address of module to set (00 to FF)
E Command to enable/disable calibration
V 1=Enable/0=Disable calibration
Response: Valid Command: !AA[CHK](cr)
Invalid Command: ?AA[CHK](cr)
Syntax or communication error may get no response.
! Delimiter for valid command
? Delimiter for invalid command
AA Address of responding module (00 to FF)
Example:
Command: $010 Receive: ?01
Perform Address 01 span calibration, returns the command is
invalid before enabling calibration.
Command: ~01E1 Receive: !01
Set Address 01 to enable calibration, return successful.
Command: $010 Receive: !01
Preform Address 01 span calibration, return successful.
Related Command:
Sec. 2.5, $AA0, Sec. 2.6, $AA1
Related Topic:
Sec. 1.9, Calibration