Section 4 Serial Controller
4.47
1. In this manner, value for
Time
will be the response time of the
instrument.
To measure the data transfer rate of a particular instrument, (the
time needed by the device to transfer
n
number of bytes), set the
trigger conditions as follows:
MATCH
= the talk address of the
instrument,
TRIG DELAY
= 1,
POST COUNT
= the larger the
number here, the more accurate the reading. In this manner, the
value for
Bytes/sec.
will be the desired information.
MODE Any
BUS STATES None
QUERY None
RESPONSE
EXAMPLE
PRINT#1,"MEASURE"
Read and display data transfer rate of
INPUT#1,A$:PRINT A$
last triggered record sequence