Measurement-computing Analyzer488 Instrukcja Użytkownika Strona 120

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 281
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 119
Section 4 Serial Controller
4.26
DISARM
The
DISARM
command prevents the Analyzer488 from sending an event status
message to the serial host, even when an
ARM
ed condition occurs. A user's program
can still check for the events by using the
STATUS
command.
If the
DISARM
command is issued without specifying any events, then all event
message reporting will be disabled.
The
ARM
command must be used to re-enable the event responses after the event
occurs.
This command does not have a query option although
ARM?
may be used to see
which events have been armed.
TYPE System Command
SYNTAX DISARM
[;][event[event]]
event
is one or more of
SRQ
,
RECORD
,
STEP
, or
TRIGGER.
RESPONSE None
MODE
Any
BUS STATES None
QUERY None
RESPONSE
EXAMPLES
PRINT#1,"DISARM"
Disable all conditions
PRINT#1,"DISARM SRQ"
Do not send '
SRQ
' string if SRQ
line is asserted.
Przeglądanie stron 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 280 281

Komentarze do niniejszej Instrukcji

Brak uwag