2.3.6 Isolated Grounds / Single-Ended Inputs
Single-ended inputs can be used to monitor isolated inputs, though the use of the differential mode will increase you
system’s noise immunity. Figure 2-9 below shows the recommended connections is this configuration
.
Figure 2-9
.
Isolated Grounds / Single-Ended Inputs
2.3.7 Isolated Grounds / Differential Inputs
Optimum performance with isolated signal sources is assured with the use of the differential input setting. Figure 2-10
below shows the recommend connections is this configuration..
Figure 2-10. Isolated Grounds / Differential Inputs
10
Isolated Signal Source
Connected to a Single-Ended Input
I/O
Connector
+
-
Input
Amp
To A/D
LL GND
CH IN
A/D Board
Isolated
signal
source
+
-
Input
Amp
To A /D
A/D Board
I/O
Connector
LL GND
CH High
CH Low
Signal Source
and A/D Board
Already Isolated.
Already isolated si
g
nal source
and A/D board connected to
a differential input.
GND
These
g
rounds are
electrically isolated.
10 K
10K is a recommended value. You may short LL GND to CH Low
instead, but this will reduce your systems noise immunity.