The following sections depicts recommended input wiring schemes for each of the seven possible input
configuration/grounding combinations.
NOTE: For simplicity, the input multiplexers are not shown in the following diagrams.
2.3.1
Common Ground / Single-Ended Inputs
Single-ended is the recommended configuration for common ground connections. However, if some of your inputs are
common ground and some are not, we recommend you use the differential mode. There is no performance penalty (other
than loss of channels) for using a differential input to measure a common ground signal source. However the reverse is
not true. Figure 2-5 below shows a basic connection diagram for a common ground / single-ended input system.
Figure 2-5
.
Common Ground / Single-Ended Input
2.3.2 Common Ground / Differential Inputs
The use of differential inputs to monitor a signal source with a common ground is a acceptable configuration though it
requires more wiring and offers half the channels of a single-ended configuration. Figure 2-6 below shows the basic
connections in this configuration.
Figure 2-6. Common Ground / Differential Inputs
8
+
-
Input
Amp
To A /D
A/D Board
I/O
Connector
LL GND
CH IN
Signal
Source with
Comm on Gnd
Optional wire
since signal source
and A/D board share
common ground
Si
g
nal source and A/D board
sharin
g
common
g
round connected
to sin
g
le-ended input.
+
-
Input
Amp
To A /D
A/D Board
I/O
Connector
LL GND
CH High
CH Low
Signal
Source with
Common Gnd
Optional wire
since signal source
and A/D board share
common ground
Required connection
of LL GND to CH Low
Si
g
nal source and A/D board
sharin
g
common
g
round connected
to differential input.