3.3.1 Common Ground / Single-Ended Inputs
Single-ended is the recommended configuration for common ground connections. However, if some of your inputs are
common ground and some are not, we recommend you use the differential mode. There is no performance penalty (other
than loss of channels) for using a differential input to measure a common ground signal source. However the reverse is
not true. Figure 3-4 below shows a recommended connection diagram for a common ground / single-ended input system
.
NOTE: For simplicity, the input multiplexer is not shown in the following diagrams.
Figure 3-4.
Single-Ended
Input
3.3.2 Common Ground / Differential Inputs
The use of differential inputs to monitor a signal source with a common ground is a acceptable configuration though it
requires more wiring and offers fewer channels than selecting a single-ended configuration. Figure 3-5 shows the
recommended connections in this configuration.
Figure 3-5. Common Ground - Differential Input
14
+
-
Input
Amp
To A /D
A/D Board
I/O
Connector
LL GND
CH IN
Signal
Source w ith
Common G nd
Optional wire
since signal source
and A/D board share
common ground
Si
g
nal source and A/D board
sharin
g
common
g
round connected
to sin
g
le-ended input.
+
-
Input
Amp
To A /D
A/D Board
I/O
Connector
LL GND
CH High
CH Low
Signal
Source with
Comm on Gnd
Optional wire
since signal source
and A/D board share
common ground
Required connection
of LL GND to CH Low
Si
g
nal source and A/D board
sharin
g
common
g
round connected
to differential input.
Przeglądanie stron 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 35 36

Komentarze do niniejszej Instrukcji

Brak uwag