Measurement-computing PCM-DAS16D/12 Instrukcja Użytkownika Strona 19

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 36
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 18
4.2WIRING CONFIGURATIONS
Combining all the grounding and input type possibilities provides us with the follow-
ing potential connection configurations. The combinations along with our recommen-
dations on usage are shown in the chart below.
RecommendedDifferential Inputs
Already Isolated
Grounds
AcceptableSingle-ended Inputs
Already Isolated
Grounds
Unacceptable without
adding Isolation
Differential Inputs
Common Mode
Voltage > +/-10V
Unacceptable without
adding Isolation
Single-Ended Inputs
Common Mode
Voltage > +/- 10V
RecommendedDifferential Inputs
Common Mode
Voltage < +/-10V
Not RecommendedSingle-Ended Inputs
Common Mode
Voltage < +/-10V
AcceptableDifferential InputsCommon Ground
RecommendedSingle-Ended InputsCommon Ground
OUR VIEW
INPUT
CONFIGURATION
GROUND
CATEGORY
The following sections depicts recommended input wiring schemes for each of the 8
possible input configuration/grounding combinations.
4.2.1 Common Ground / Single-Ended Inputs
Single-ended is the recommended configuration for common ground connections.
However, if some of your inputs are common ground and some are not, we recom-
mend you use the differential mode. There is no performance penalty (other than loss
of channels) for using a differential input to measure a common ground signal source.
15
Przeglądanie stron 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 35 36

Komentarze do niniejszej Instrukcji

Brak uwag