eZ-NDT 938691 Investigation Mode Testing 2-1
Investigation Mode Testing 2
1. Gather known “good” and “bad” parts …2-1
2. Set up the analyzer, hammer, microphone, and computer …… 2-1
3. Launch eZ-NDT …… 2-2
4. Open a New eZ-NDT Project …… 2-3
5. Select Investigation Mode …… 2-4
6. Configure the Analyzer …… 2-4
7. Select Plot Display Type …… 2-4
8. Turn on Data Acquisition ……2-4
9. Test All Good Parts …2-4
10. Test All Bad Parts ……2-6
11. Compare the Results of Good and Bad Tests ……2-7
12. Configure the Limits …… 2-7
13. Verifying the Limits and Ranges …… 2-9
The following process can be used to determine whether eZ-NDT testing is feasible for the part that is
being tested. Typically this is an iterative process in which the user:
establishes the resonant signature of the part
determines the distinct peaks of the signature
uses the peaks to set up eZ-NDT test criteria
1. Gather known “good” and “bad” parts.
Locate both good and bad parts. Typically 10 of each are sufficient. Make sure the Bad parts include all
possible reject (failure) modes. The good parts will be used to test for NDT feasibility.
Block Diagram of Typical NDT Setup