Table 5-5. 82C55 Registers
Configure 82C55 None. No read back on 82C55BASE + 403
Port C OutputPort C InputBASE + 402
Port B OutputPort B InputBASE + 401
Port A OutputPort A Input of 82C55 BASE + 400
WRITE FUNCTIONREAD FUNCTIONADDRESS
The two groups of ports, group A and group B, may be independently programmed in one of several modes. The most
commonly used mode is mode 0, input / output mode. The codes for programming the 82C55 in this mode are shown
in Table 5-6 below.
Table 5-6. 82C55 Configuration Coding
ININININ1559B1111
OUTINININ1549A0111
INOUTININ153991011
OUTOUTININ152980011
ININOUTIN147931101
OUTINOUTIN146920101
INOUTOUTIN145911001
OUTOUTOUTIN144900001
INININOUT1398B1110
OUTININOUT1388A0110
INOUTINOUT137891010
OUTOUTINOUT136880010
ININOUTOUT131831100
OUTINOUTOUT130820100
INOUTOUTOUT129811000
OUTOUTOUTOUT128800000
CLBCUADECHEXD0D1D3D4
NOTE: D7 is always 1. D6, D5, and D2 are always 0.
6 CALIBRATION AND TEST
Every board is fully tested and calibrated before leaving the factory. For normal environments a calibration interval of
six months to one year is recommended. If frequent variations in temperature or humidity are common then
recalibrate at least once every three months. It takes less than 30 minutes to calibrate the board.
You will not need an extender card to calibrate theboard but you will need to have the cover off your computer with
the power on, so trim pots can be adjusted during calibration.
The A/D is calibrated by applying a known voltage to an analog input channel and adjusting trim pots for offset and
gain. There are two trim pots requiring adjustment to calibrate the analog input section. The entire procedure is
described in detail in the InstaCal
TM
, calibration routine.
The CIO-DAS16/M1/16 should be calibrated for the range you intend to use it in. When the range is changed, slight
variations in zero and full scale may result. These variations can be measured and removed in software if necessary.
16