Worst Case Temperature Measuerment Error vs. Personal Daq3000 Ambient Temperature
with Thermocouple at 0ºC (Excludes Thermocouple Error);
AutoZero Disabled; Selected Thermocouple Types: T, J, K, and E
0
1
2
3
4
5
6
7
8
-30 -20 -10 0 10 20 30 40 50 60 70
Ambient Temperature (ºC)
Temperature Measurement Error (ºC) for Types T, J, K, E
J
K
T
E
Worst Case Temperature Measuerment Error vs. PDaq3000 Ambient Temperature
with Thermocouple at 0ºC (Excludes Thermocouple Error);
AutoZero Enabled; Selected Thermocouple Types: T, J, K, and E
0.0
0.5
1.0
1.5
2.0
2.5
3.0
-30 -20 -10 0 10 20 30 40 50 60 70
Ambient Temperature (ºC)
Temperature Measurement Error (ºC) for Types T, J, K, E
J
K
T
E
Note: Specifications are subject to change without notice.