Measurement-computing WBK40 Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Sprzęt komputerowy Measurement-computing WBK40. Measurement Computing WBK40 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 34
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
WBK40 and WBK41
Thermocouple & Multi-Function I/O Modules
For use with WaveBook/516E and WBK25
WBK40
14 Channel Thermocouple Input Module,
Front & Rear Views
Overview ….. 1
What are WBK40 and WBK41? …… 1
Front Panel Connectors …… 2
Rear Panel Connectors ….. 4
Setup …… 5
Hardware Setup …… 5
Software Setup …… 14
Operational Aspects ...... 14
Block Diagrams …… 14
Thermocouple Measurements …… 15
Digital I/O, WBK41 Only …… 16
Synchronous Input Operations …… 16
Output Channel Configuration …… 17
Asynchronous I/O Operations …… 18
Analog Outputs, WBK41 Only (DBK46 Required) …… 19
Using a Temperature Calibrator …… 20
CE Compliance …… 21
Fuse Replacement …… 23
P1 - DB37 Connector Pinout …… 26
P2 – DB37 Connector Pinout (WBK41 Only) …… 27
Specifications …… 28
WBK40 & WBK41 Installation Guide
(p/n 1066-0901)
WBK41
Multi-Function Thermocouple I/O Module,
Front & Rear Views
Overview
What are WBK40 and WBK41?
The WBK40 is a 14-Channel Thermocouple Input Module. The WBK41 is a Multi-Functional I/O Module. The
latter includes the thermocouple functionality of the WBK40 and additional features, as will be discussed shortly.
Both the WBK40 and WBK41 attach to any one of the three DB25 parallel expansion ports on a WaveBook/516E
or WBK25. Because the modules each have a built-in 16 bit, 200-kHz A/D converter, measurements do not
consume valuable sampling time from the host WaveBook’s 1-MHz A/D converter. For this reason, there are no
speed implications to WaveBook [or other WBK] measurements when a WBK40 or a WBK41 is attached.
Measurements can be made synchronous with the WaveBook/516E, providing precise time correlation between
readings from both measurement devices.
Features common to WBK40 and WBK41
Provides a means of adding from 14 to 224 thermocouple inputs to the WaveBook/516E or WBK25
Provides linearized and cold-junction compensated readings for all thermocouple types
Exhibits 1.0°C [or better] accuracy for most ranges and TC types
Consumes no measurement bandwidth from the WaveBook/516E
Detects open thermocouples
Additional features of the WBK41
16 Digital I/O Lines (via front panel screw-terminal blocks)
24 Digital I/O Lines (via a rear panel 37-pin connector, P2)
4 Counter Inputs (via front panel screw-terminal blocks)
2 Timer Outputs (via front panel screw-terminal blocks)
4 Channel, 16-bit, 100 kHz Analog Output (via front panel screw-terminal
blocks; but only when a factory-installed DBK46 option is in place)
Thermocouple & Multifunctional Modules
926896
WBK40 & WBK41, pg. 1
Przeglądanie stron 0
1 2 3 4 5 6 ... 33 34

Podsumowanie treści

Strona 1 - Overview

WBK40 and WBK41 Thermocouple & Multi-Function I/O Modules For use with WaveBook/516E and WBK25 WBK40 14 Channel Thermocouple Inpu

Strona 2 - Front Panel Connectors

What are “T-Cables” and how are they used? CAUTION Turn power OFF to the acquisition devices and to externally connected equipment prior to conne

Strona 3

Connecting to the Device Interface Side – WBK41 System Examples WARNING Electrical Shock Hazard! To avoid possible injury and equipment damage, tu

Strona 4 - Rear Panel Connectors

Channel Input Side Device Interface Side A WaveBook/516A with a WBK25, WBK41, and DBK84 Note 1 DBK84 WB

Strona 5

Setting “Unique” Channel Addresses for DBK84 Modules All DBK84 modules that are connected to WBK40 or WBK41 systems must have unique channel address

Strona 6

Software Setup You will need to use the Daq* Configuration control panel applet to add the WBK40 or WBK41 device to the parent (first level) device,

Strona 7

WBK40 Block Diagram Thermocouple Measurements Each WBK40 and WBK41 module’s internal circuitry includes cold junction compensation (CJC) for direct

Strona 8

In regard to adding thermocouples to WBK40, WBK41, or DBK84 modules, Channels 1 through 7 are used for the first seven thermocouples, and channels 9 t

Strona 9

Counter Input Channels The devices allow synchronous scanning of the 4 16-bit counter input channels. The four 16-bit counter channels can also be ca

Strona 10 - CAUTION

Asynchronous I/O Operations - WBK41 The WBK41 allows asynchronous input of any counter or digital channel that is not currently configured for sync

Strona 11 - WARNING

Analog Output Channels (The DBK46 Factory Install Option is Required) When a WBK41 has a DBK46 option installed, the module has the ability to output

Strona 12

Front Panel Connectors WBK41 Front Panel WBK41, Front Panel Connectors Note that the WBK40’s Front Panel is identical, except it has no Screw-Termina

Strona 13

DBK46 is installed at the factory. To verify that a DBK46 is installed, simply check the acquisition software’s Analog Output Window for the presence

Strona 14 - Operational Aspects

CE Compliance Overview CE-compliant products bear the “CE” mark and include a Declaration of Conformity stating the particular specifications and con

Strona 15 - Thermocouple Measurements

For clarification, terms used in some Declarations of Conformity include: • pollution degree: any addition of foreign matter, solid, liquid or gaseo

Strona 16 - Digital Input Channels

CE Enhancements for DBKs The following CE enhancements are described in the individual document modules of the DBK Cards and Modules User’s Manual (p

Strona 17 - WBK40 & WBK41, pg. 17

You will need the following: Phillips Screwdriver • • • • • 3AG Fuse Puller, or needle-nose pliers* 3/16-inch hex-nut driver Grounding wrist strap and

Strona 18

5. Locate fuse F201 (see the preceding figure). 6. While wearing a grounded wrist strap, remove and replace the bad fuse. Ensure that the new fuse

Strona 19

P1 DB37 Connector Pinout Analog I/O Pin Signal Name Description for P1 Pin Use 1 +5 VDC Ex

Strona 20

P2 DB37 Connector Pinout (WBK41 Only) Digital I/O P2 does not apply to WBK40. Pin Signal Name Description for P2 Pin

Strona 21 - CE Compliance

Specifications – WBK40 and WBK41 Power Supply Voltage Range: 10 to 30 VDC Power Out, from P1: +15 VDC @ 500 mA (via pin 21) –15 VDC @ 500 mA (v

Strona 22 - Safety Conditions

TC Accuracy at Measurement Temperature in °C (±°C) Type Min Max -100 0 100 300 500 700 900 1100 1400 J -200 760 0.6 0.5 0.5 0.6 0.7 0.7

Strona 23 - Fuse Replacement

WBK40 Front Panel WBK40 Front Panel Connectors The Mini-TC and Analog Common Connectors exist on both the WBK40 and WBK41 Modules. Item Descripti

Strona 24

Triggering WBK40, Trigger Sources Four, individually selectable for starting and stopping an acquisition. Stop acquisition can occur on a different ch

Strona 25 - WBK40 & WBK41, pg. 25

Additional Specifications for WBK41 In addition to the preceding specifications, the following apply to WBK41. They do not apply to WBK40. Power Out

Strona 26

Front Panel – Digital I/O Characteristics: Note: The following characteristics apply to the16 Digital I/O lines that connect to the front panel’s rem

Strona 27

Ordering Information Modules 14-channel thermocouple input module for the WaveBook/516E and WBK25 WBK40 Multi-function I/O module for the WaveBook/5

Strona 28

pg. 34, WBK40 & WBK41 926896 Thermocouple and Multifunctional Modules

Strona 29 - Input Sequencer

Rear Panel Connectors WBK41, Rear Panel Connectors Note that WBK40’s Rear Panel is identical, except it has no P2 connector. Page 1 includes a

Strona 30 - Triggering

Setup Reference Note For a quick set up involving connecting a WBK40 or a WBK41 to a WaveBook/516E or to a WBK25, refer to the WBK40 & WBK41 Inst

Strona 31 - WBK40 & WBK41, pg. 31

DIGITAL I/O (2 Terminal Blocks) POWER [Out] COUNTER - TMR D/A (DAC0, DAC1, DAC2, DAC3) WBK41, Front

Strona 32 - Frequency/Pulse Generators

Open Thermocouple Detection The WBK40, WBK41, and DBK84 modules are equipped with open thermocouple detection for each channel. This means that a bro

Strona 33 - Ordering Information

Connecting to the Device Interface Side – WBK40 System Examples WARNING Electrical Shock Hazard! To avoid possible injury and equipment damage, turn

Strona 34

DBK84 DBK84 DBK84 WBK40 WaveBook/516E WBK17 WBK10A Note 1 Channel Input Side

Powiązane modele: WBK41

Komentarze do niniejszej Instrukcji

Brak uwag