DAQFlex Message-based Firmware Specification Analog input calibration messages
13
OFFSET
Set the offset in hexadecimal bytes for the current AICAL:RANGE setting.
Message
"AICAL{ch}:OFFSET/HEX=hex value"
or
"AICAL{ch}:OFFSET=0xhex value"
Refer to the device-specific information in the Hardware Reference chapter to see
whether "/HEX" is supported.
Response
"AICAL{ch}:OFFSET"
ch
The channel number.
hex value
The calibration offset in hexadecimal bytes.
Example
"AICAL{0}:OFFSET/HEX=0x84B575AF"
Note
The hex value is used to preserve the precision of the double precision floating point
value.
Get the offset in hexadecimal bytes for the current AICAL:RANGE setting.
Message
"?AICAL{ch}:OFFSET/HEX"
or
"?AICAL{ch}:OFFSET"
Refer to the device-specific information in the Hardware Reference chapter to see
whether "/HEX" is supported.
Response
"AICAL{ch}:OFFSET=0xhex value"
ch
The channel number.
hex value
The calibration offset in hexadecimal bytes.
RANGE
Set the range to calibrate.
Message
"AICAL:RANGE=value"
Response
"AICAL:RANGE"
value
The range value.
Example
"AICAL:RANGE=BIP10V"
Get the range to calibrate.
Message
"?AICAL:RANGE"
Response
"AICAL:RANGE=value"
value
The range value.