E-1608 Ethernet-based High-speed Multifunction DAQ Document Revision 1 December 2014 © Copyright 2014 User's Guide
E-1608 User's Guide Installing the E-1608 10 1. Assuming you have successfully connected to a local network, determine the IP address of the de
11 Chapter 3 Functional Details Analog input modes The E-1608 can acquire analog input data in two basic modes – software paced and hardware paced. S
E-1608 User's Guide Functional Details 12 The maximum communication distance without using a repeater is 100 meters. You can send your data 100
E-1608 User's Guide Functional Details 13 The E-1608 pinout locations are shown in Figure 3 below. Figure 3. E-1608 pinout Signal connections
E-1608 User's Guide Functional Details 14 Leave unused input channels either floating or tied to an AGND terminal. Source impedances should be k
E-1608 User's Guide Functional Details 15 Analog output Two 16-bit analog outputs are available at AOUT0 and AOUT1. Each analog output channel
E-1608 User's Guide Functional Details 16 Figure 7 shows the jumper configured for pull-up and pull-down. Caution! The discharge of static elec
E-1608 User's Guide Functional Details 17 Mechanical drawings Figure 8. E-1608 device circuit board dimensions
E-1608 User's Guide Functional Details 18 Figure 9. E-1608 bottom enclosure dimensions
E-1608 User's Guide Functional Details 19 Figure 10. E-1608 top enclosure dimensions
HM E-1608.docx Your new Measurement Computing product comes with a fantastic extra — Management committed to your satisfaction! Thank you for choo
20 Chapter 4 Specifications All specifications are subject to change without notice. Typical for 25 °C unless otherwise specified. Specifications in
E-1608 User's Guide Specifications 21 Accuracy Analog input DC voltage measurement accuracy Table 2. DC Accuracy components and specifications.
E-1608 User's Guide Specifications 22 Parameter Condition Specification Output coupling DC Power on and reset state DACs cleared to uncalibrat
E-1608 User's Guide Specifications 23 Parameter Specification Input high voltage threshold 2.0 V min Input high voltage limit 5.5 V absolute max
E-1608 User's Guide Specifications 24 Parameter Specification Schmitt trigger hysteresis 1.01 V typ 0.6 V min 1.5 V max Input high voltage thres
E-1608 User's Guide Specifications 25 Power Table 15. Power specifications Parameter Condition Specification External power supply 5V, 1A Suppl
E-1608 User's Guide Specifications 26 Network factory default settings Table 18. Factory default specifications Parameter Specification Factory
E-1608 User's Guide Specifications 27 Screw terminal connector Table 23. Screw terminal connector specifications Parameter Specification Connect
Declaration of Conformity According to ISO/IEC 17050-1:2010 Manufacturer: Measurement Computing Corporation Address: 10 Commerce Way Suite 100
Measurement Computing Corporation 10 Commerce Way Suite 1008 Norton, Massachusetts 02766 (508) 946-5100 Fax: (508) 946-9500 E-mail: [email protected]
3 Table of Contents Preface About this User's Guide ...
E-1608 User's Guide 4 Settling time ...
5 Preface About this User's Guide What you will learn from this user's guide This user's guide describes the Measurement Computing E-1
6 Chapter 1 Introducing the E-1608 The E-1608 is compatible with both TCP/IP (IPv4 only) and user datagram protocol (UDP) network protocols. The E-1
E-1608 User's Guide Introducing the E-1608 7 Functional block diagram E-1608 functions are illustrated in the block diagram shown here. Figure
8 Chapter 2 Installing the E-1608 Unpacking As with any electronic device, you should take care while handling to avoid damage from static electrici
E-1608 User's Guide Installing the E-1608 9 Configuring network settings The following E-1608 network settings are software-selectable. Only one