pg. 4, WBK12A and WBK13A
988396
Programmable Low-Pass Filter Cards
Accuracy and Noise Specifications
WaveBook/516 Series
(Alone)
WaveBook/516 Series
with a WBK12A or a WBK13A
(Note 3)
Voltage
Range
Accuracy
(Note 2)
One Year, 18-28
°C
Input Noise
LSB rms
DC-500KHz
(typical)
(Note 4)
Accuracy
(Note 2)
One Year, 18-28
°C
Input Noise
LSB rms
(typical)
±
%
reading
±
%
range
±
%
reading
±
%
range
1KHz
Filter
Filter
Bypass
0 to +10V .012% .008% 2 .012% .008% 2.2 2.2
0 to +5V (10A)
0 to +4V
(
516
)
.012% .009% 2 .012% .009% 2.2 2.2
0 to +2V .012% .012% 3 .012% .012% 2.2 3
0 to +1V
(
10A onl
y)
.012% .018% 3 .012% .018% 2.2 3
0 to +.5V .018% .033% 2.2 6
0 to +.2V .018% .08% 2.2 12
0 to +.1V .018% .16% 2.2 20
-10 to +10V .012% .008% 2 .012% .008% 2.2 2.2
-5 to +5V .012% .008% 2 .012% .008% 2.2 2.2
-2 to +2V .012% .009% 2 .012% .009% 2.2 3
-1 to +1V .018% .012% 3 .018% .012% 2.2 3.3
-.5 to +.5V
(10A only)
.018% .018% 5 .018% .018% 2.2 6
-.2 to +.2V .018% .033% 2.2 12
-.1 to +.1V .018% .08% 2.2 20
-.05 to +.05V
(
10A onl
y)
.018% .16% 440
Notes
: 1. Specifications assume differential input scan, unfiltered
2. Accuracy specification is exclusive of noise.
3. Unipolar ranges are unavailable for a WaveBook/516 Series with a
WBK11A, WBK12A, or WBK13A option installed.
Unipolar mode is available with WBK10A and any option.
4. Maximum limit is 1.3X typical.