WavePort User’s Manual
09-2900
Specifications 10-4
WavePort/V Specifications
IOtech Option Function Internal/External
WBK11A 8-Channel Simultaneous Sample and Hold Internal
WBK12A 8-Channel Programmable Low-pass Filter Internal
WBK13A 8-Channel Programmable Low-pass Filter with
Simultaneous Sample and Hold
Internal
WBK61, 62 High Voltage Adapters and Probes External
This section provides Analog Specifications Applicable to:
WavePort/V8 - WaveBook (8 channels total)
WavePort/V16 - WaveBook with one expansion module (16 channels total)
WavePort/V24 - WaveBook with two expansion modules (24 channels total)
Input Voltage Ranges (DC Specifications):
Standard Unit With WBK11A
(Note 3)
With WBK12A/13A
(Note 3)
Voltage
Range
Accuracy
(Note 2)
One Year, 18-28
°C
Input Noise
LSB rms
dc-500KHz
(typical)
Accuracy
(Note 2)
One Year, 18-28
°C
Input Noise
LSB rms
dc-500KHz
(typical)
Accuracy
(Note 2)
One Year, 18-28
°C
Input Noise
LSB rms
(typical)
±
%
reading
±
%
range
±
%
reading
±
%
range
±
%
reading
±
%
range
1KHz
Filter
Filter
Bypass
0 to +10V .012% .008% 2 .012% .008% 2 .012% .008% 2.2 2.2
0 to +5V (10A)
0 to +4V
(
516
)
.012% .009% 2 .012% .009% 2 .012% .009% 2.2 2.2
0 to +2V .012% .012% 3 .012% .012% 3 .012% .012% 2.2 3
0 to +1V
(
10A onl
y)
.012% .018% 3 .012% .018% 3 .012% .018% 2.2 3
0 to +.5V N/A N/A N/A .018% .033% 6 .018% .033% 2.2 6
0 to +.2V N/A N/A N/A .018% .08% 8 .018% .08% 2.2 12
0 to +.1V N/A N/A N/A .018% .16% 15 .018% .16% 2.2 20
-10 to +10V .012% .008% 2 .012% .008% 2 .012% .008% 2.2 2.2
-5 to +5V .012% .008% 2 .012% .008% 2 .012% .008% 2.2 2.2
-2 to +2V .012% .009% 2 .012% .009% 2 .012% .009% 2.2 3
-1 to +1V .018% .012% 3 .018% .012% 3 .018% .012% 2.2 3.3
-.5 to +.5V
(10A only)
.018% .018% 5 .018% .018% 6 .018% .018% 2.2 6
-.2 to +.2V N/A N/A N/A .018% .033% 8 .018% .033% 2.2 12
-.1 to +.1V N/A N/A N/A .018% .08% 15 .018% .08% 2.2 20
-.05 to +.05V
(
10A onl
y)
N/A N/A N/A .018% .16% 26 .018% .16% 440
Notes: 1. Specifications assume differential input scan, unfiltered
2. Accuracy specification is exclusive of noise.
3. Unipolar ranges unavailable for WaveBook with WBK11A, WBK12A, or WBK13A options installed.
Available with Module 1 and Module 2 with any option.