USB-3110 User's Guide Specifications
19
Note 3: The maximum differential non-linearity specification applies to the entire 0 to 50 °C temperature
range of the USB-3110. This specification also accounts for the maximum errors due to the
software calibration algorithm (in Calibrated mode only) and the DAC8554 digital to analog
converter non-linearities.
Note 4: The USB-3110 voltage outputs should not be kept in a short-circuit condition for longer than the
specified limit of 100 ms. For those applications that may potentially exceed the 40 mA
maximum current limit or the 100 ms short-circuit condition, external current limiting must be
used to prevent potential damage to the USB-3110.
Table 2. Absolute accuracy specifications – calibrated output, VOUTx Rload = 20 mA fixed resistive load
Table 3. Absolute accuracy components specifications – calibrated output
Absolute accuracy at FS (±mV)
Table 4. Relative accuracy specifications
Analog output calibration
Table 5. Analog output calibration specifications
On-board precision reference
DC level: 5.000 V ±1 mV max.
Long term stability: ±10 ppm/SQRT(1000 hrs)
Voutx Rload = 20 mA fixed
resistive load