USB-3102 User's Guide Specifications
21
Note 2: The USB-3102 output voltage level defaults to 0V whenever the output voltage range is
reconfigured.
The USB-3102 output voltage level will also default to 0V:
1) Whenever the host PC is reset, shut down or suspended.
2) If a reset command is issued to the device.
Note 3: The duration of this particular output transient is highly dependent on the enumeration process of
the host PC. Typically the output of the USB-3102 is stable after 2 seconds.
Note 4: The maximum differential non-linearity specification applies to the entire 0 to 50 °C temperature
range of the USB-3102. This specification also accounts for the maximum errors due to the
software calibration algorithm (in Calibrated mode only) and the DAC8554 digital to analog
converter non-linearities.
Table 2. Absolute accuracy specifications – calibrated output
Table 3. Absolute accuracy components specifications – calibrated output
Absolute accuracy at FS (±mV)
Table 4. Relative accuracy specifications
Analog current output
Table 5. Analog current output specifications
Number of channels
(Note 5)
Compliance voltage range
(Note 6)
Differential non-linearity (Note 7)
±1.25 LSB typ.
-2 LSB to +1 LSB max.
±0.25 LSB typ.
-1 LSB to +1 LSB max.
Compliance voltage = +13V
Resistive load = 100Ω
±0.05% of full scale range
All zeros written to IOUTx
channel
100nA typ. (each individual IOUT
channel)
Full scale step to 1LSB accuracy
Compliance voltage = 12V,
Rload = 500Ω
100 Hz max., system dependent
100 Hz/#ch max., system dependent
Komentarze do niniejszej Instrukcji