USB-2408 Series User's Guide Specifications
26
The multiple-channel throughput rate is calculated using this formula:
Maximum throughput = , where n is the number of channels
Table 12. Multiple-channel throughput rate specifications (Hz)
Analog voltage output (USB-2408-2AO only)
Unused AOUTx output channels should be left disconnected.
The USB-2408-2AO output voltage level defaults to 0 V whenever the host PC is reset, shut down or
suspended, or if a reset command is issued to the device.
The outputs may have a transient during startup. The duration of the output transient depends highly on the
enumeration process of the host PC. Typically, the output of the USB-2408-2AO is stable after two seconds.
Table 13. USB-2408-2AO analog voltage output specifications
Digital-to-analog
converter
±10.05 V (software-selectable)
Host PC is reset, powered on, suspended or a reset
command is issued to device
Duration: 2 s
Amplitude: 2 V p-p
Duration: 50 ms
Amplitude: 5 V peak
Differential non-linearity
Output short-circuit
protection