1-2 Introduction and Quick Start
04-26-01
TempBook User’s Manual
TempBook related Software – What you should install from the Data Acquisition CD
For 32-Bit Support
*
For 16-Bit Support
*
32-bit TempBook Support*
is used with Windows 9x, ME, NT, and 2000
Operating Systems and installs:
DaqView (32-bit)
DaqViewXL (32-bit)
File Converter (32-bit)
DaqCal (32-bit)
32-bit drivers for VB, C, and Delphi
PDF version of Programmer’s Manual
16-bit TempBook Support*
is used with Windows 3.x Operating
Systems and installs:
TempView (16-bit)
PostView (16-bit)
TempCal (16-bit)
16-bit drivers for VB, C, PASCAL, and
DOS
DIAdem Post Acquisition Data Analysis
Program
Acrobat Reader
Acrobat Reader
*32-bit refers to support, not to A/D resolution. *16-bit refers to support, not to A/D resolution.
Available Accessories
Additional accessories that can be ordered for the TempBook/66 include:
WBK20A PCMCIA interface card and cable
DBK30A Rechargeable battery module
WBK21 ISA Interface Card
Specifications
General
Connector
: Connects to a PC via an included parallel port cable; user signals are connected via screw terminals on a
removable screw-terminal card.
Thermocouple Types
: J, K, S, T, E, B, R & N
Input Ranges
:
Unipolar
: .05, 0.1, 0.2, 0.5, 1, 2, 5, 10V
Bipolar
: ±0.025, .05, 0.1, 0.25, 0.5, 1, 2.5, 5 V
Analog Inputs
: 8 differential or 16 single-ended voltage or 8 differential thermocouple inputs
Analog-to-Digital Converter
: 12-bit with no missing codes, unsigned binary output format.
Cold Junction Sensor Output
: 100 mV/°C
Input Impedance
: 100K/100M Ohm, Switch-selectable on a per-channel basis in parallel with switched 100 pF
Input RC Filter -3dB Frequency
: 15.9 kHz switch-selectable on a per channel basis
Gain Accuracy
: 0.1%
Maximum Input Voltage
: ±15 V
CMRR (Input Stage)
: 90 dB typ, DC to 60 Hz
Offset
: Software-compensated
Offset Drift
: Software-compensated
Channel Sequencer
Depth
: 512 locations
Speed
: 10 µs per channel, fixed
Interval Between Scans
: 10 µs to 10 hours, software-programmable
Gains
: Sequencer programmable on a per-channel basis.
Unipolar/Bipolar
: Sequencer programmable on a per channel basis.
Single-Ended/Differential: Software programmable for all channels