C-4 TEDS 928693 Appendix C
The Basic TEDS contains the manufacturer’s identification, model identification, version letter, and the serial
number, which are common to all types of sensors. The Standard TEDS, however, contains data unique to the
sensor’s principle of operation. For instance, a piezoelectric accelerometer differs from a strain gage load cell, so the
Standard TEDS EEPROM contains some common and some unique information. Both sensor TEDS list Calibration
Date, Measurement Range, and Electrical Output, but they differ in as many as 12 other parameters. The
accelerometer has 6 parameters that the strain gage lacks, and the strain gage has 6 parameters the accelerometer
lacks. The User Area contains information such as Sensor Location, Calibration Due Date, and Calibration Table.
TEDS Configuration: Accelerometer
Basic TEDS Manufacturer ID 25
Model ID 5226
Version Letter A
Serial Number 11521S
Standard TEDS Calibration Date Jan. 6, 2004
(Subtemplate) Sensitivity @ Reference 1.12E+03 mV/g
Reference Frequency 110.0 Hz
Reference Temperature 25˚C
Measurement Range ±50 g
Electrical Output ±5 V
Quality Factor
Temperature Coefficient -0.27%/˚C
Direction (X, Y, Z) X
User Area Sensor Location Matrix A: B-3
Calibration Due Date Jun. 15, 2004
Fig. C.04. Table 2. Example for an Accelerometer.
TEDS Configuration: Strain-Gage Load Ce
ll
Basic TEDS Manufacturer ID 56
Model ID 350A
Version Letter E
Serial Number 150-00878T
Standard TEDS Calibration Date Mar. 15, 2004
Measurement Range ±250 lb
Electrical Output ±5 mV/V
Bridge Impedance 500
Measurement Range ±75 g
Excitation, Nominal 24 Vdc
Excitation, Minimum 10 Vdc
Excitation, Maximum
Response Time
Calibration TEDS Calibration Table -2.06 mV/V, -150 lb
-1.02 mV/V, -75 lb
0.03 mV/V, 0 lb
0.95 mV/V, 75 lb
1.89 mV/V, 150 lb
User Area Sensor Location L23-3
Calibration Due Date Jun. 15, 2004
Fig. C.04. Table 3. Example for a Strain-Gage Load Cell.