USER’S MANUAL
DaqBook/2000 Series and
DaqOEM/2000 Series
16-Bit, 200 kHz Multifunction Data Acquisition Systems
DaqBook/2001
DaqBook/2005
DaqBook/2020
DaqOEM/2001
DaqOEM/2005
Requires one of the following
Operating Systems:
Windows XP
Windows Vista
DaqBook/2000 Series
&
DaqOEM/2000 Series
1121-0901 rev 3.0
*372335A-01*
372335A-01
Measurement Computing
10 Commerce Way
Norton, MA 02766
(508) 946-5100
Fax: (508) 946-9500
www.mccdaq.com
Przeglądanie stron 0
1 2 3 4 5 6 ... 89 90

Podsumowanie treści

Strona 1 - DaqOEM/2000 Series

USER’S MANUAL DaqBook/2000 Series and DaqOEM/2000 Series 16-Bit, 200 kHz Multifunction Data Acquisition Systems DaqBook/2001DaqBook/2005DaqBook/202

Strona 2

Step 4 - Configure Computer Network Settings Note: We recommend that you discuss this procedure with your Network Administrator before proceeding. 1.

Strona 3

Daq Systems 1 Daq Systems, the Modular Concept …… 1-1 DaqBooks, DaqBoards, and Daq PC-Cards ….. 1-2 Using DBK Cards and Modules for Signal Conditio

Strona 4 - 1086-0922

1-2 Daq Systems 958795 DaqBook/2000 Series and DaqOEM/2000 Series DaqView and DASYLab can only be used with one DaqBook/2000 Series, DaqOEM S

Strona 5 - Manual Layout

DaqBook/2000 Series and DaqOEM/2000 Series 958795 Daq Systems 1-3 Daq Data Acquisition Devices Category Device Description DaqBook/2000 Series

Strona 6

1-4 Daq Systems 958795 DaqBook/2000 Series and DaqOEM/2000 Series Using DBK Cards and Modules for Signal Conditioning The DBK signal-cond

Strona 7 - Table of Contents

Daq Software The Daq devices have software options capable of handling most applications. Three types of software are available: • ready-to-use gra

Strona 8

1-6 Daq Systems 958795 DaqBook/2000 Series and DaqOEM/2000 Series

Strona 9 - Quick Start

DaqBook/2000 Series and DaqOEM/2000 Series 2 Device Overviews I/O Comparison Matrix …… 2-1 Overview …… 2-2 Product Features …… 2

Strona 10 - *324631A-01*

2-2 Device Overviews 898095 DaqBook/2000 Series and DaqOEM/2000 Series Overview DaqBook/2020, /2001, /2005, DaqOEM/2001, and DaqOEM/2005 are

Strona 11 - Daq Systems 1

Triggering DaqBook/2020, /2001, /2005, DaqOEM/2001, and DaqOEM/2005 each support a full complement of trigger modes including: • Hardware analog trig

Strona 13 - Daq Data Acquisition Devices

2-4 Device Overviews 898095 DaqBook/2000 Series and DaqOEM/2000 Series Digital Input Channels The devices allow for either synchronous scannin

Strona 14

DaqBook/2000 Series and DaqOEM/2000 Series 898095 Device Overviews 2-5 Asynchronous I/O Operations The DaqBook/2000 Series and DaqOEM/2000 Ser

Strona 15 - Daq Software

2-6 Device Overviews 898095 DaqBook/2000 Series and DaqOEM/2000 Series Synchronous Operation Matrix The following table applies to DaqBook/20

Strona 16

DaqBook/2001 & DaqOEM/2001 16 Analog Input via P1 4 Analog Output via P3 40 Digital I/O via P2 & P3 4 Counter In 2 Timer Out via P3 Su

Strona 17 - Device Overviews

2-8 Device Overviews 898095 DaqBook/2000 Series and DaqOEM/2000 Series DaqBook/2005 & DaqOEM/2005 16 Analog Input via P1 No DAC Chann

Strona 18 - Product Features

DaqBook/2000 Series and DaqOEM/2000 Series 898095 Device Overviews 2-9 DaqBook/2020’s on-board scan sequencer lets you select up to 16384 channe

Strona 19 - Channel Sampling Interval

2-10 Device Overviews 898095 DaqBook/2000 Series and DaqOEM/2000 Series

Strona 20

Block Diagrams 3 DaqBook/2001, DaqOEM/2001, DaqBook/2005, & DaqOEM/2005 …… 3-1 DaqBook/2020 …… 3-2 Detail “TC” (Thermocouple Input Circuitry) f

Strona 21 - Counter Input Channels

3-2 Block Diagrams 966895 DaqBook/2000 Series and DaqOEM/2000 Series DaqBook/2020 Block Diagram DaqBook/2020 Block Diagram

Strona 22 - DaqOEM/2005

Detail “TC” (Thermocouple Input Circuitry) for DaqBook/2020 Detail “V” (Differential Voltage Input Circuitry) for DaqBook/2020 DaqBook/2000 Ser

Strona 23

Warranty Information Contact Measurement Computing by phone, fax, or e-mail in regard to warranty-related issues: Phone: (508) 946-5100, fax: (508) 9

Strona 24 - DaqBook/2020

3-4 Block Diagrams 966895 DaqBook/2000 Series and DaqOEM/2000 Series

Strona 25

Connectors, Indicators, Cables, & Pinouts 4 DaqBook/2020 …… 4-1 DaqBook/2001 and DaqBook/2005 …… 4-3 DaqOEM/2001 and Daq

Strona 26

Rear Panel The DaqBook/2020 rear panel contains P1 and P2 DB37 connectors, a MAC Address Label, DIN5 power connector, power switch, 2 Synchronizatio

Strona 27 - Block Diagrams 3

DaqBook/2001 and DaqBook/2005 DaqBook/2001 and /2005 provide for up to 256 channels of analog input expansion. They have 40 built-in digital I/O cha

Strona 28 - DaqBook/2020 Block Diagram

POWER Switch: A rocker-type switch with a “0” label for Power Off, and a “1” for Power On. POWER IN: +10 to +30 VDC, through a socket type DIN5 conne

Strona 29

Cables Should you need help in regard to cables or have any other product inquiries please contact the factory. Cables for use with DaqBook/2000 Ser

Strona 30

4-6 Connectors, Indicators, Cables, & Pinouts 898095 DaqBook/2000 Series and DaqOEM/2000 Series

Strona 31

Pinouts P1 and JP1 – Analog I/O …… 4-8 P2 and JP2 – Digital I/O …… 4-9 P3 and JP3 – Pulse/Frequency/Digital I/O (Does not apply to DaqBook/2020E)…

Strona 32

P1 and JP1 Analog I/O P1, DB37 Connector for DaqBook/2000 Series Devices JP1,

Strona 33

P2 and JP2 Digital I/O P2, DB37 Connector

Strona 34 - DaqOEM/2001 and DaqOEM/2005

iv 898695 DaqBook/2000 Series and DaqOEM/2000 Series Your order was carefully inspected prior to shipment. When you receive your system, car

Strona 35

P3 and JP3 Pulse/Frequency/Digital I/O JP3, 40-pin header for DaqOEM/2000

Strona 36

DaqBook/2000 and DaqOEM/2000 Series User’s Manual 959295 Software 5-1 Software 5 Overview …… 5-1 Out-of-the-BoxTM Software …… 5-1 Drivers for Th

Strona 37 - Pinouts

5-2 Software 959295 DaqBook/2000 and DaqOEM/2000 Series User’s Manual Reference Notes: ¾ The software document modules, DaqView, DaqViewXL, an

Strona 38 - This notch is

Configuring the System for Operation 6 Step 1 - Install the Software ….. 6-2 Step 2 - Determine the type of Network C

Strona 39

CAUTION Turn off power to the system devices and externally connected equipment before connecting cables. Electric shock or damage to equipment c

Strona 40

Dedicated Network - with a direct cable connection from the PC to the device In this scenario a DaqBook/2020 is connected directly to an Ethernet jack

Strona 41 - Software 5

LAN with a DHCP Server (Local Area Network with a Dynamic Host Configuration Protocol server) Many corporations use the LAN/Server with DHCP arrang

Strona 42 - Language Drivers

STEP 3 – Connect the System Components Reference Note: For information regarding connections, refer to Chapter 4, Connectors, Indicators, Cables,

Strona 43

STEP 4 – Power-up the System What you will need: A power supply with a range of +10 VDC to +30 VDC. The power supply needs to have a male DIN5 co

Strona 44 - Install the Software

STEP 5 – Configure the Computer’s Network Settings [Applies to “dedicated networks” only] If using a LAN (Local Area Network), which has a DHCP s

Strona 45

DaqBook/2000 Series and DaqOEM/2000 Series 898695 v PDF 457-0905 DBK Options.pdf The DBK Option Ca

Strona 46

Local Area Connection Properties 5. Double-click the “Internet Protocol (TCP/IP)” component (previous figure). The “Internet Protocol (TCP/IP) Pro

Strona 47

Configure the Computer’s TCP/IP settings as follows. Internet Protocol (TCP/IP) Properties 6. Select the “Use the following IP Address” radio

Strona 48 - Power-up the System

STEP 6 – Configure Device Network Settings using DaqIPConfig Applies only to a LAN (Local Area Network), which has a no DHCP server. If using a LA

Strona 49

STEP 7 – Configure and test the System using the Daq* Configuration Applet The Daq* Configuration applet, designed for 32-bit Windows NT/2000/XP s

Strona 50

3. Set the properties of the first-level device. In this step you will set the device properties according to one of the following two methods, depe

Strona 51

4. Test the system connections. a. Make sure the device has been properly installed and is powered-on. b. Make sure all cables are properly and s

Strona 52

Reference Notes: Depending on your application and system setup, related information can be found in the following sources: ¾ The DBK Options Manu

Strona 53

CE Compliance & Noise Considerations 7 Overview ……7-1 CE Standards and Directives …… 7-1 Safety Conditions ……7-2 Emissions/Immunity Conditions

Strona 54 - For DEDICATED Networks

For clarification, terms used in some Declarations of Conformity include: • pollution degree: any addition of foreign matter, solid, liquid or gaseo

Strona 55

CE Enhancements for DBKs The following CE enhancements are detailed in the DBK Cards and Modules User’s Manual (p/n 457-0905). • DBK41/CE • Edge G

Strona 56

vi 898695 DaqBook/2000 Series and DaqOEM/2000 Series This page is intentionally blank.

Strona 57 - CE Standards and Directives

Noise Considerations Controlling electrical noise is imperative because it can present problems even with the best measurement equipment. Most labora

Strona 58 - Safety Conditions

Ethernet CE Kit Installation Instructions Introduction The Ethernet CE Kit includes three ferrite clamp-on inductive collars

Strona 59 - CE Enhancements for DBKs

Install the Ferrite Inductive Collars 1. Place one ferrite collar at each end of the Ethernet cable. Snap the two collars shut. See photos. 2.

Strona 60

Taking Measurements 8 Introduction …… 8-1 Scanning …… 8-1 Triggering …… 8-2 Synchronous I/O Operations…… 8-3 Asynchronous I/O Operations …… 8-7 Usi

Strona 61 - Connect the Ground Line

8-2 Taking Measurements 898695 DaqBook/2000 and DaqOEM/2000 Series Triggering Triggering can be the most critical aspect of acquiring data. Da

Strona 62

Pre- and Post-Triggering Modes. Six modes of pre- and post-triggering are supported, providing a wide variety of options to accommodate any measuremen

Strona 63 - Taking Measurements 8

8-4 Taking Measurements 898695 DaqBook/2000 and DaqOEM/2000 Series Synchronization Model for DaqBook/2000 and DaqOEM/2000 Series Devices, D

Strona 64 - Triggering

SYNC Notes: • When a DaqBook is programmed as a synchronization slave, the TTL trigger source is automatically derived from the SYNC port. • Whe

Strona 65 - Synchronous I/O Operations

8-6 Taking Measurements 898695 DaqBook/2000 and DaqOEM/2000 Series Digital I/O Channels There are 24 Digital I/O channels available to DaqBook

Strona 66

DaqBook/2000 and DaqOEM/2000 Series 898695 Taking Measurements 8-7 Asynchronous I/O Operations Each DaqBook allows asynchronous input of an

Strona 67

DaqBook/2000 Series and DaqOEM/2000 Series 999895 vii Table of Contents Quick Start 1 – Daq Systems 2 – DaqBook/2000 Series and DaqOEM/2000 Series

Strona 68 - (AO0, AO1, AO2, AO3)

8-8 Taking Measurements 898695 DaqBook/2000 and DaqOEM/2000 Series Counter Input Channels (CTR0, CTR1, CTR2, CTR3) DaqBooks and DaqOEMs have

Strona 69

DaqBook/2000 and DaqOEM/2000 Series 959095 Calibration 9-1 Calibration

Strona 70 - (CTR0, CTR1, CTR2, CTR3)

9-2 Calibration 959095 DaqBook/2000 and DaqOEM/2000 Series

Strona 71 - Using DaqCal.exe

Troubleshooting and Customer Support 10 Electrostatic Discharge (ESD), Handling Notice…… 10-1 Product Care …… 10-1 ReadMe Files and the Install CD-

Strona 72

10-2 Troubleshooting & Customer Support 897995 DaqBook/2000 and DaqOEM/2000 - IO version ReadMe Files and the Install CD-ROM The Install CD-R

Strona 73 - Product Care

DaqBook/2000 and DaqOEM/2000 Series – IO version 897995 Troubleshooting & Customer Support 10-3 Ethernet Problems • • • • • Make sure the

Strona 74 - DaqCOM Issues

10-4 Troubleshooting & Customer Support 897995 DaqBook/2000 and DaqOEM/2000 - IO version Frequently Asked Questions (1) Topic: Environment

Strona 75 - Ethernet Problems

DaqBook/2000 and DaqOEM/2000 Series – IO version 897995 Troubleshooting & Customer Support 10-5 Customer Support If you want to Expand or

Strona 76 - Frequently Asked Questions

10-6 Troubleshooting & Customer Support 897995 DaqBook/2000 and DaqOEM/2000 - IO version

Strona 77

DaqBook/2000 and DaqOEM/2000 Series 899595 Specifications 11-1 Specifications 11 DaqBook/2020, /2001, /2005, DaqOEM/2001, and DaqOEM/200

Strona 78

viii 999895 DaqBook/2000 Series and DaqOEM/2000 Series This page is intentionally blank.

Strona 79 - Specifications 11

Specifications are subject to change without notice. 11-2 Specifications 958893 DaqBook/2000 and DaqOEM/2000 User’s Manual Settling Time: 5 µ

Strona 80

Specifications are subject to change without notice. DaqBook/2000 and DaqOEM/2000 Series 898695 Specifications 11-3 Thermocouple Input Specif

Strona 81 - TC Accuracy

Specifications are subject to change without notice. 11-4 Specifications 898695 DaqBook/2000 and DaqOEM/2000 User’s Manual Input Sequencer Ana

Strona 82 - External Sync Ports

Specifications are subject to change without notice. DaqBook/2000 and DaqOEM/2000 Series 898695 Specifications 11-5 5. Counter/Totalizer Trig

Strona 83 - Digital I/O

Specifications are subject to change without notice. 11-6 Specifications 898695 DaqBook/2000 and DaqOEM/2000 User’s Manual Frequency/Pulse Cou

Strona 84 - Frequency/Pulse Generators

Specifications are subject to change without notice. DaqBook/2000 and DaqOEM/2000 Series 898695 Specifications 11-7 Accessories and Cables

Strona 85 - Accessories and Cables

Specifications are subject to change without notice. 11-8 Specifications 898695 DaqBook/2000 and DaqOEM/2000 User’s Manual Manuals Manual S

Strona 86

Glossary 959395 G-1 Glossary Acquisition A collection of scans acquired at a specified rate as controlled by the sequencer. Analog A signal of va

Strona 87 - Glossary

G-2 959395 Glossary Digital-to-Analog Converter (DAC) A circuit or device that converts digital values (binary bits), into

Strona 88 - 959395 Glossary

Glossary 959395 G-3 Trigger An event to start a scan or mark an instant during an acquisition. The event can be defined in various ways; e.g., a TTL

Strona 89 - Glossary 959395 G-3

1121-0940, rev 1.0 324631A-01 Printed in Hungary DaqBook & DaqOEM

Strona 90

G-4 959395 Glossary

Powiązane modele: DaqOEM 2000 Series