4 CALIBRATION AND TEST
The CIO-DAS08-AOx is supplied with InstaCal software which also performs calibration and board testing.
Every board is fully tested and calibrated before shipment. For normal environments, a calibration interval of 6 months to one
year is recommended. If frequent variations in temperature or humidity are common, re-calibrate at least every three months. It
requires less than 30 minutes to calibrate the CIO-DAS08-AOx.
The CIO-DAS08-AOx does not require recalibration when moved from one computer to another. The reason is that the board has
a stable DC/DC converter on-board which supplies the analog +/−15V voltages.